Visual Field Testing with Head-Mounted Perimeter 'imo'

Published

2016

Audience

Therapeutic Area

Content Type

AI Generated Summary*

A portable head-mounted perimeter, imo, usable without a darkened room, was piloted on 40 eyes of 20 patients with glaucoma. Alongside monocular testing, it can present stimuli unpredictably to either eye without occlusion. Testing followed a 30-2 grid, with 4-2 dB thresholding on imo and SITA Standard on the Humphrey Field Analyzer (HFA); mean sensitivity did not differ significantly across the three conditions (P = 0.99), and correlations ranged from r = 0.94 to 0.98. Differences between the two imo modes averaged 0.21 dB (right) and -0.11 dB (left). The binocular mode took 17.3 minutes versus 16.0 for the HFA, and no participant identified the tested eye. Repeatability was not examined, and the manufacturer funded it.

*This summary was generated by AI and is published unedited. Oku does not alter these summaries. It may contain errors or omissions and is provided for general informational purposes only. Oku does not guarantee its accuracy, completeness, or reliability. For authoritative information, please refer to the original, peer-reviewed article.