The development of a reference database with the Topcon 3D OCT-1 Maestro

Published

2018

Audience

Therapeutic Area

Content Type

AI Generated Summary*

Reference limits for a spectral domain OCT platform were built from 399 healthy eyes at seven US clinical sites; 504 were screened, and mean age was 46.3 years (18 to 88), 43% male. Non-crossing quantile regressions covarying age, plus disc area for optic nerve measures, generated 1st through 99th percentile limits. Averages from the 12 by 9 mm wide scan were roughly 237 μm foveal thickness, 71 μm ganglion cell plus inner plexiform layer, 106 μm ganglion cell complex and 105 μm circumpapillary nerve fiber layer, similar to 6 by 6 mm scans. Ganglion cell measures thinned with age, while disc area strongly influenced nerve fiber layer thickness. Values resembled published ranges, though the report warns against swapping devices.

*This summary was generated by AI and is published unedited. Oku does not alter these summaries. It may contain errors or omissions and is provided for general informational purposes only. Oku does not guarantee its accuracy, completeness, or reliability. For authoritative information, please refer to the original, peer-reviewed article.

At a glance

Authors
Michael Chaglasian; Murray Fingeret; Pinakin Gunvant Davey; Wei-Chieh Huang; Danny Leung; Edmund Ng; Charles A. Reisman
Publication
Clinical Ophthalmology
Keywords
3D OCT-1 Maestro, Circumpapillary Retinal Nerve Fiber Layer, Macular Measurements, Normal Eyes, Optic Disc, Spectral Domain Optical Coherence Tomography