Precision and agreement of axial length in paediatric population measured with MYAH and AL-Scan biometers
AI Generated Summary*
Axial length precision and agreement were compared for two myopia management biometers, MYAH and AL-Scan, in Spanish schoolchildren aged 7 to 10. Of 187 participants (mean age 8.5 years, mean spherical equivalent +0.22 D), subsets of 74 to 144 joined each test, on undilated right eyes. Both devices showed a Sw of 0.01 mm, with 95% of repeats expected inside 0.04 mm, under single-operator and two-operator testing alike, although AL-Scan values shifted between operators (p = 0.002). Mean between-device difference was 0.00 mm, limits of agreement -0.04 to 0.05 mm (p = 0.021). Every figure sat below the 0.10 mm threshold the authors treat as clinically meaningful, supporting their view that the instruments are interchangeable for myopia monitoring.
*This summary was generated by AI and is published unedited. Oku does not alter these summaries. It may contain errors or omissions and is provided for general informational purposes only. Oku does not guarantee its accuracy, completeness, or reliability. For authoritative information, please refer to the original, peer-reviewed article.