Clinical validation of a new optical biometer for myopia control in a healthy pediatric population
AI Generated Summary*
Fifty-one healthy Spanish children aged 5 to 18 (mean 10.8 years) each contributed one randomly chosen eye to repeatability testing of a new multifunction optical biometer: flat and steep corneal radius and white-to-white distance were captured twice, axial length three times. Within-subject standard deviations were 0.018, 0.021, 0.071 and 0.017 mm, with intraclass correlation coefficients of at least 0.971. Agreement was examined only in a 30-eye subgroup, measured once with a comparator myopia management device, which gave larger corneal radii (K1 0.013 mm; K2 0.058 mm), while the newer unit read white-to-white 0.151 mm wider and axial length 0.030 mm longer. Only the K1 difference was nonsignificant; all gaps were small, and interchangeable use was endorsed with caution.
*This summary was generated by AI and is published unedited. Oku does not alter these summaries. It may contain errors or omissions and is provided for general informational purposes only. Oku does not guarantee its accuracy, completeness, or reliability. For authoritative information, please refer to the original, peer-reviewed article.