Clinical Evaluation of a Novel Preloaded Intraocular Lens Delivery System During Routine Cataract Surgery

Published

2020

Audience

Therapeutic Area

Content Type

AI Generated Summary*

A prospective, noncomparative, open-label, manufacturer-funded study at four Australian and New Zealand centers gathered surgeon and technician ratings of a preloaded delivery system, a disposable cartridge on a reusable titanium handpiece, for implanting a one-piece monofocal lens through 2.2 to 2.4 mm incisions. Five surgeons and 14 technicians rated 91 eyes; overall performance was judged acceptable in all, with 85.7% given the top score. Surgeons estimated mean savings of 54.9 seconds per case versus non-preloaded systems, although one site reported little benefit. Every lens was centered and correctly folded, and no ocular adverse events, haptic breakage, or lens marking were reported through the 1-day postoperative visit. Limitations noted include no control arm and reliance on questionnaires.

*This summary was generated by AI and is published unedited. Oku does not alter these summaries. It may contain errors or omissions and is provided for general informational purposes only. Oku does not guarantee its accuracy, completeness, or reliability. For authoritative information, please refer to the original, peer-reviewed article.

At a glance

Authors
Daniel Black; Dean Corbett; Timothy V. Roberts; Brendan Cronin; Pamela J. Smith; D. Priya Janakiraman; Beth E. Jackson
Publication
Clinical Ophthalmology
Keywords
IOL, Cataract Surgery, Clinical Handleability, Preloaded Delivery System
Year
2020
View primary source

This is the evidence

See where we take it next

Join the Network