A Comparison of Optical Biometers Used in Children for Myopia Control

Published

2023

Audience

Therapeutic Area

Content Type

AI Generated Summary*

Axial length and corneal readings from four biometers were compared retrospectively in 22 myopic children (44 eyes; mean age about 11 years, spherical equivalent -3.53 D), with the IOLMaster 700 as reference and 16 children retested on at least one device. Intra-subject repeatability for axial length was 0.05 mm with the reference, 0.06 mm for Myopia Master and Myah, and 0.04 mm for Lenstar, implying waits of 5.0 to 6.7 months before yearly growth of 0.1 mm could be detected reliably. Lenstar read 0.02 mm longer and Myopia Master keratometry 0.21 D lower; J0 differences reached significance but stayed under 0.10 D. Agreement was described as good overall, with at least six months advised between repeat axial checks.

*This summary was generated by AI and is published unedited. Oku does not alter these summaries. It may contain errors or omissions and is provided for general informational purposes only. Oku does not guarantee its accuracy, completeness, or reliability. For authoritative information, please refer to the original, peer-reviewed article.

At a glance

Authors
Ann-Isabel Mattern; Kai Neller; Machteld Devenijn; Hartmut Schwahn; Achim Langenbucher; Berthold Seitz; Hakan Kaymak
Publication
Klinische Monatsblatter Fur Augenheilkunde
Keywords
Anterior Chamber, Axial Length, Eye, Biometry, Cornea, Myopia